Colored Glass Filters
COLORED GLASS FILTERS
Multiband Calibration Filters (V, HY)
V10 and V11 contain neodymium, and HY1, holmium. All three exhibit outstanding absorption at specific wavelengths.
The filters are used for wavelength calibration and V10 as a dual-color-type filter.
Application: Wavelength calibration
Images: HY1 / V11
Transmittance
Multiband Calibration Filter: Data
Item | Standard Value | Color specification (Standard illuminants A) | Thickness | Refractive Index | Chemical property | Thermal properties | Mechanical properties | Specific gravity | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
T550(%) | T586(%) | x | y | mm | 600nm | DW | DA | Tg | Ts | α -30/70 | α 100/300 | HK | FA | d | |
V10 | >80 | <2 | 0.447 | 0.386 | 2.5 | 1.530 | 4 | 1 | 500 | 545 | 103 | 117 | 520 | 130 | 2.87 |
V11 | - | <1 | - | - | 2.0 | 1.541 | 1 | 1 | 596 | 657 | - | 87.57 | 555 | 110 | 2.83 |
HY1 | >80 | <5 | 0.431 | 0.411 | 2.5 | 1.509 | 4 | 1 | 590 | 640 | 102 | 116 | 520 | 130 | 2.72 |
* The listed data are standard values. As we are always improving our products, the data listed are subject to change without notice.